Specialists in Electrochemistry & Materials, Energy & Power

Sensit Smart

Smartphone Potentiostat

  • Capable of EIS up to 200 kHz
  • Built around the EmStat Pico module
  • Potential range -1.7 to +2 V
  • Current ranges 100 nA – 5 mA (max ±3 mA)
 

Description:

The Sensit Smart is the smallest potentiostat available. You can directly connect the Sensit Smart toa smart phone or tablet and can be controlled via an Android app PStouch. You can alos connect it to your PC using a USB-C female to USB-A cable and then control it using the software PSTrace.

 

The Sensit Smart supports most common electrochemical techniques, including Cyclic Voltammetry, Square Wave Voltammetry and Impedance Spectroscopy (FRA/EIS).

 

Sensit Smart comes with-

  • Dummy Cell (SPE Version)
  • PSE to screw-terminal adapter
  • USB-C Female to USB-A cable
  • USB-C Female to Micro USB adapter
  • USB-C port protector
  • PSTrace software for Windows (activation code for my.palmsens.com)
  • PStouch app for Android (find it in the Google Play Store)

Techniques:

Voltammetric Techniques:

  • Linear Sweep Voltammetry (LSV)
  • Cyclic Voltammetry (CV)

Pulsed Techniques:

  • Differential Pulse Voltammetry (DPV)
  • Square Wave Voltammetry (SWV)
  • Normal Pulse Voltammetry (NPV)

Amperometric Techniques:

  • Chronoamperometry (CA)
  • Chronocoulometry (CC)
  • MultiStep Amperometry (MA)
  • Pulsed Amperometric Detection (PAD)

Galvanotstatic Techniques:

  • Open Circuit Potentiometry (OCP)

Electrochemical Impedance Spectroscopy (EIS):

  • Potential scan
  • Fixed potential
  • Time scan

Specifications:

GeneralLow Speed ModeHigh Speed ModeMax Range Mode
Full dc-potential range-1.2 to +2 V-1.7 to +2 V-1.7 to +2 V
Dynamic dc-potential range [1]2.2 V1.2 V 2.6 V
Compliance voltage -2.0 to +2.3 V [2] -2.0 to +2.3 V [2] -2.0 to +2.3 V [2]
Maximum current± 3 mA
± 3 mA
± 3 mA
Max. acquisition rate
(datapoints/s)
100 1000 100
Supports FRA/EIS NoYesNo

 

Potentiostat (controlled potential mode)
Channels1 (1x WE, 1x RE and 1x CE)
Applied dc-potential resolution537 µV395 µV932 µV
Applied potential accuracy< 0.2%< 0.5%< 0.5%
Current ranges100 nA, 2 uA,
4 uA, 8 uA,
16 uA, 32 uA,
63 uA, 125 uA,
250 uA, 500 uA,
1 mA, 5 mA
100 nA, 1 uA,
6 uA, 13 uA,
25 uA, 50 uA,
100 uA,
200 uA, 1 mA,
5 mA
100 nA, 1 uA,
6 uA, 13 uA,
25 uA, 50 uA,
100 uA,
200 uA, 1 mA,
5 mA
Current accuracy< 0.5 % of the selected
current range
< 2% for 100 nA current range
< 1% of the selected current range< 1% of the selected current range
Measured current resolution 0.006% of selected current range (5.5 pA on 100 nA range)
Measured potential resolution56 uV

 

FRA / EIS (impedance measurements)
Frequency range0.016 Hz to 200 kHz
Ac-amplitude range1 mV to 0.25 V rms, or 0.708 V p-p

 

FRA / EIS (impedance measurements)
Frequency range0.016 Hz to 200 kHz
Ac-amplitude range1 mV to 0.25 V rms, or 0.708 V p-p

 

Instrument
Power and communicationUSB-C
Storage4000 datapoints on-board
Size43 x 25 x 11 mm (excl. USB connector)
Weight10 g
BatteryNo
Auxiliary portNo
Sensor connectorSPE (pitch: 2.54 mm)
Operation temperature range0 °C to +40 °C