Specialists in Electrochemistry & Materials, Energy & Power

Coin Cell Holder

Fully isolated V+/- and I+/- leads. Can accommodate standard coin cell sizes.

Novonix cell holders are designed to be used with NOVONIX 2A Channel Modules. They carry a circuit board with an integrated RTD temperature sensor and a header to optionally connect an external RTD for cell surface measurements*. They can accommodate Coin, Standard Cylindrical (18650,26650,21700,4680), and Pouch cells.

Standard format cell holders have maximum current ratings of 10A.

Source

CURRENT

Range: 200µA, 2mA, 20mA, 200mA, 2A
Resolution: <20ppm FSR
Accuracy: 0.01% FSR
Noise: 0.002% FSR
Temperature Coefficient: 0.002% FSR over 23 ±5°C

VOLTAGE

Range: 250mV, 5V
Resolution: 1µV
Accuracy: 100µV (<250mV), 200µV (>250mV)
Noise: 100µV
Temperature Coefficient: 0.002% FSR over 23 ±5°C

Measurement

CURRENT

Resolution: <20ppm
Accuracy: 0.005% FSR
Noise: 0.002% FSR
Temperature Coefficient: 0.002% FSR over 23 ±5°C

VOLTAGE

Resolution: <50nV on 0V-0.25V, 1µV on 0.25V-5V
Accuracy: 100µV (<250mV), 200µV (>250mV)
Noise: 100µV
Temperature Coefficient: 0.002% FSR over 23 ±5°C

Timing

Resolution: 10ms
Measurement Frequency: 6Hz
Accuracy: 2ppm

Integrated RTD Temperature Sensing

Resolution: ±0.01°C
Accuracy: ±0.5°C
Noise: ±0.05°C

Technical Considerations*

Dimensions WxHxD (In): 19” x 5” x 24”
Rack Space (U): 3U
Weight (lbs): 28
Number of Test Channels: 8
Bench and Rack Mountable?: Yes
Power Draw (VA): 320VA

Voltage: 120VAC-230VAC 50Hz or 60Hz
Minimum Discharge Voltage: 0.005V
Required Lab Temperature Range: 18°C – 28°C
Required Lab Temperature Stability: ±1°C