IviCycle - High channel count Cycler with EIS per channel.
- Multi-channel battery test system
- Multipotentiostat system – 5 electrodes inc Reference Electrodes.
- C030: 8 channels, each channel: ±30mA / 10V
- C200: 8 channels, each channel: ±200mA / ±10V
- FRA/EIS: 10µHz-20kHz on each channel for simultaneous testing.
- 19″ rack-mountable housing.
- Internal uDataSecure – experiment can run without PC.
- Optional temperature input per channel.
Description:
Dedicated battery test system
IviCycle is designed to be your perfect partner for high channel-count battery applications. It offers a low cost-per-channel, high volume testing solution and integral extensive data storage capability
Complete solution
The IviCycle is a multi-channel test system with optional impedance capability. The channels are divided over 4 x 8 channel modules which can be a mix of the different IviCycle modules to create the desired number of channels with varying compliance.
The C200 modules each have 8 channels of ±200mA/±10V compliance, the C030 have ±200mA/±10V. Modules are available either as a DC version (no EIS), or as an EIS version with impedance capability on each of the channels. This means that simultaneous/parallel impedance measurements are possible on all channels. The IviCycle has an integrated DataSecure module that stores all data for each channel completely independently from the computer. This ensures that if there is a communication loss or a computer crash/update, the measurement will continue and data will not be lost and so gives peace-of-mind for long duration, high throughput testing. The IviCycle unit is built into a 19inch rack mountable housing for easy incorporation into existing storage racks, or as a stand-alone bench-top multi-channel system.
Safety
The maximum current or potential of the instrument can be defined by the operator. Thus valuable samples may be protected, and unsafe situations prevented.
Data security & Connection
The IviCycle is equipped with an integrated DataSecure module:
- Data is stored on board independent of the computer.
- Connection of multiple PCs is possible
Software
The IviCycle is controlled via Ivium’s own software, IviumSoft. This versatile, yet intuitive software, allows instrument control, data management and analysis, etc. The IviCycle can also be controlled from LabVIEWTM, C++, Delphi, etc. A full suite of software is included as standard with each instrument.
- Control of separate channels individually
- All channels simultaneously
- Synchronous start
- Data from different channels plotted individually or in 1 single gra
Software:
The IviCycle can be controlled using Iviums own software (IviumSoft). The software allows you to control the instrument, data management and the analysis of the data collected. If you choose to use other software it can controlled using LabVIEWTM, C++, Delphi etc. The IviCycle comes with a full software which is included with each instrument.
- Control of separate channels individually
- All channels simultaneously
- Synchronous start
- Data from different channels plotted individually or in 1 single graph
Specifications:
Performance | |
---|---|
Current compliance | ±30mA |
Maximum output voltage | ±10V |
Electrode connections | 4; WE, CE, RE, S with 4mm banana plugs |
Potentiostat bandwidth | >500kHz |
Stability settings | High Speed, Standard and High Stability |
Signal acquisition | Dual channel 16bit ADC, 100,000 samples/s |
Channel combination | No |
Potentiostat | |
---|---|
Applied potential range | ±10V, 0.333mV res. |
Applied potential accuracy | 0.2% or 1mV |
Current ranges | ±10nA to ±100mA |
Measured current resolution | 0.015% of current range, minimum 1pA |
Measured current accuracy | 0.2% |
Galvanostat | |
---|---|
Applied current resolution | 0.033% of applied current range |
Applied current accuracy | 0.2% |
Potential ranges | ±10mV, ±100mV, ±1V, ±10V |
Galvanostatic current ranges | ±10µA to ±100mA |
Measured potential resolution | 0.003% of potential range, minimum 400nV |
Measured potential accuracy | 0.2% or 1mV |
Impedance Analyser | |
---|---|
Frequency range | 10µHz to 20kHz |
Amplitude | 0.15mV to 1.0V, or 0.03% to 100% of current range |
DC offset | 16bit DC offset subtraction and 2 DC-decoupling filters |
Electrometer | |
---|---|
Input impedance | >1000Gohm //<20pF |
Input bias current | <20pA |
Bandwidth | >5MHz |
Connections | |
---|---|
Analog in/out | 16bit analog input |
Instrument | |
---|---|
Interfacing | USB/LAN (Ethernet) |
Size (w x d x h) | 44 x 26 x 22cm |
Weight | 12.5kg |
Integrated DataSecure | Data acquisition time: 2ms minimum, Stored no. of data points: 20M each channel |